  • \( \newcommand{\vecs}[1]{\overset { \scriptstyle \rightharpoonup} {\mathbf{#1}} } \) \( \newcommand{\vecd}[1]{\overset{-\!-\!\rightharpoonup}{\vphantom{a}\smash {#1}}} \)\(\newcommand{\id}{\mathrm{id}}\) \( \newcommand{\Span}{\mathrm{span}}\) \( \newcommand{\kernel}{\mathrm{null}\,}\) \( \newcommand{\range}{\mathrm{range}\,}\) \( \newcommand{\RealPart}{\mathrm{Re}}\) \( \newcommand{\ImaginaryPart}{\mathrm{Im}}\) \( \newcommand{\Argument}{\mathrm{Arg}}\) \( \newcommand{\norm}[1]{\| #1 \|}\) \( \newcommand{\inner}[2]{\langle #1, #2 \rangle}\) \( \newcommand{\Span}{\mathrm{span}}\) \(\newcommand{\id}{\mathrm{id}}\) \( \newcommand{\Span}{\mathrm{span}}\) \( \newcommand{\kernel}{\mathrm{null}\,}\) \( \newcommand{\range}{\mathrm{range}\,}\) \( \newcommand{\RealPart}{\mathrm{Re}}\) \( \newcommand{\ImaginaryPart}{\mathrm{Im}}\) \( \newcommand{\Argument}{\mathrm{Arg}}\) \( \newcommand{\norm}[1]{\| #1 \|}\) \( \newcommand{\inner}[2]{\langle #1, #2 \rangle}\) \( \newcommand{\Span}{\mathrm{span}}\)\(\newcommand{\AA}{\unicode[.8,0]{x212B}}\)

    This page describes a double beam UV-visible absorption spectrometer.

    The overall design

    If you pass white light through a colored substance, some of the light gets absorbed. A solution containing hydrated copper(II) ions, for example, looks pale blue because the solution absorbs light from the red end of the spectrum. The remaining wavelengths in the light combine in the eye and brain to give the appearance of cyan (pale blue).

    Some colorless substances also absorb light - but in the ultra-violet region. Since we can't see UV light, we don't notice this absorption. Different substances absorb different wavelengths of light, and this can be used to help to identify the substance - the presence of particular metal ions, for example, or of particular functional groups in organic compounds.

    The amount of absorption is also dependent on the concentration of the substance if it is in solution. Measurement of the amount of absorption can be used to find concentrations of very dilute solutions. An absorption spectrometer measures the way that the light absorbed by a compound varies across the UV and visible spectrum.

    A simple double beam spectrometer

    We'll start with the full diagram, and then explain exactly what is going on at each stage.

    spectrometer.gif

    The light source

    You need a light source which gives the entire visible spectrum plus the near ultra-violet so that you are covering the range from about 200 nm to about 800 nm. (This extends slightly into the near infra-red as well.)

    You can't get this range of wavelengths from a single lamp, and so a combination of two is used - a deuterium lamp for the UV part of the spectrum, and a tungsten / halogen lamp for the visible part. The combined output of these two bulbs is focused on to a diffraction grating.

    The diffraction grating and the slit

    You are probably familiar with the way that a prism splits light into its component colors. A diffraction grating does the same job, but more efficiently.

    slit.gif

    The blue arrows show the way the various wavelengths of the light are sent off in different directions. The slit only allows light of a very narrow range of wavelengths through into the rest of the spectrometer. By gradually rotating the diffraction grating, you can allow light from the whole spectrum (a tiny part of the range at a time) through into the rest of the instrument.

    The rotating disks

    This is the clever bit! Each disk is made up of a number of different segments. Those in the machine we are describing have three different sections - other designs may have a different number.

    spinningdisc.gif

    The light coming from the diffraction grating and slit will hit the rotating disk and one of three things can happen.

    1. redpath.gif
    2. If the original beam of light from the slit hits the mirrored section of the first rotating disk, it is bounced down along the green path. After the mirror, it passes through a reference cell (more about that later). Finally the light gets to the second disk which is rotating in such a way that it meets the transparent section. It goes straight through to the detector.
    3. If the light meets the first disk at the black section, it is blocked - and for a very short while no light passes through the spectrometer. This just allows the computer to make allowance for any current generated by the detector in the absence of any light.

    The sample and reference cells

    These are small rectangular glass or quartz containers. They are often designed so that the light beam travels a distance of 1 cm through the contents. The sample cell contains a solution of the substance you are testing - usually very dilute. The solvent is chosen so that it doesn't absorb any significant amount of light in the wavelength range we are interested in (200 - 800 nm). The reference cell just contains the pure solvent.

    The detector and computer

    The detector converts the incoming light into a current. The higher the current, the greater the intensity of the light. For each wavelength of light passing through the spectrometer, the intensity of the light passing through the reference cell is measured. This is usually referred to as Io - that's I for Intensity.

    The intensity of the light passing through the sample cell is also measured for that wavelength - given the symbol, I.

    If I is less than Io, then obviously the sample has absorbed some of the light. A simple bit of math is then done in the computer to convert this into something called the absorbance of the sample - given the symbol, A.

    For reasons which will become clearer when we do a bit of theory on another page, the relationship between A and the two intensities is given by:

    \[A=\log \left( \dfrac{I}{I_o} \right) \]

    On most of the diagrams you will come across, the absorbance ranges from 0 to 1, but it can go higher than that. An absorbance of 0 at some wavelength means that no light of that particular wavelength has been absorbed. The intensities of the sample and reference beam are both the same, so the ratio Io/I is 1. Log10 of 1 is zero.

    An absorbance of 1 happens when 90% of the light at that wavelength has been absorbed - which means that the intensity is 10% of what it would otherwise be.

    In that case, Io/I is 100/I0 (=10) and log10 of 10 is 1.

    The absoprtion spectrum is a plot of absorbance against wavelength. The output might look like this:

    absgraph1.gif

    This particular substance has what are known as absorbance peaks at 255 and 395 nm. How these arise and how they are interpreted are diskussed on another page.


    This page titled A Double Beam Absorption Spectrometer is shared under a CC BY-NC 4.0 license and was authored, remixed, and/or curated by Jim Clark.